OECDReference EN/AU/PS1105248
Accu-Chek Solo Cannula Assembly & Pump Holder and Accu-Chek Solo Supply Kit
Official information organised for easier consultation. The source publication remains authoritative.
Identification
Product identification
- Product
- Accu-Chek Solo Cannula Assembly & Pump Holder and Accu-Chek Solo Supply Kit
Official information
Risk and measures
- Risk or hazard
- Insulin leakage has occurred with the Accu-Chek Solo Cannula Assembly and Pump Holder. If the insulin leakage is not immediately noticed when using the Accu-Chek Solo system this could potentially lead to an under-delivery of insulin. If insulin is under-delivered, this may lead to adverse health events including severe hyperglycaemia or the user may experience diabetic ketoacidosis (DKA) and may require hospitalisation or intervention from a medical professional.
- Recommendations
- Consumers should check whether they have an affected product and if so stop using them. Consumers should contact Roche Diabetes Customer Care to make arrangements to receive a replacement. For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01541-1 For further information, please contact Roche Diabetes Customer Care by phone on 1800 251 516.
Official information
Economic operators
- Manufacturer
- Roche Diabetes Care Australia Pty Limited
Territory
Geography
- Distribution
- Retirada en Australia
- Country of recall
- Australia