Invacare Perfecto2 V Oxygen Concentrator

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Invacare Perfecto2 V Oxygen Concentrator

Official information

Risk and measures

Risk or hazard
The sound abatement washer may break down resulting in metal on metal wear inside the Pressure Equalisation (PE) valve. This may lead to high temperatures and overheating, separation of the shroud from the device, and loud noise, and can result in a fire that will shut down the device. If the device shuts down this can cause hypoxemia (low oxygen in the blood) from "loss of use". There is an increased risk of fire, which may result in serious injury and property damage.
Recommendations
Consumers should check the model and serial number of their machine to see if their device is affected. Invacare Australia will contact affected users in writing with instructions on how to to make arrangements to replace the PE valve assembly. The replacement of the PE valve assembly will be done within 6 months. In the interim, consumers should continue to use the device in accordance with its labelling/user manual. Labelling that accompanies the device instructs the user to always have a backup source of oxygen readily available. For further information, consumers can contact Invacare Australia on 1800 460 460.

Official information

Economic operators

Manufacturer
Invacare Australia Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia