Invacare Perfecto2 V Oxygen Concentrator

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Invacare Perfecto2 V Oxygen Concentrator

Información oficial

Riesgo y medidas

Riesgo o peligro
The sound abatement washer may break down resulting in metal on metal wear inside the Pressure Equalisation (PE) valve. This may lead to high temperatures and overheating, separation of the shroud from the device, and loud noise, and can result in a fire that will shut down the device. If the device shuts down this can cause hypoxemia (low oxygen in the blood) from "loss of use". There is an increased risk of fire, which may result in serious injury and property damage.
Recomendaciones
Consumers should check the model and serial number of their machine to see if their device is affected. Invacare Australia will contact affected users in writing with instructions on how to to make arrangements to replace the PE valve assembly. The replacement of the PE valve assembly will be done within 6 months. In the interim, consumers should continue to use the device in accordance with its labelling/user manual. Labelling that accompanies the device instructs the user to always have a backup source of oxygen readily available. For further information, consumers can contact Invacare Australia on 1800 460 460.

Información oficial

Operadores económicos

Fabricante
Invacare Australia Pty Ltd

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia