Bock Nursing Care Beds - Battery

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Bock Nursing Care Beds - Battery

Official information

Risk and measures

Risk or hazard
In beds with double drives, manufactured in the period from January 2012 to May 2013, some batteries may contain diodes which may be defective. Failure of the diode may lead to damage of the emergency lowering battery. If the bed cannot be lowered this may lead to injury.
Recommendations
Consumers with impacted stock should contact Alphacare to organise correction. Alphacare will arrange a technician to visit and make the necessary corrections. For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/