OECDReferencia EN/AU/PS1080123
Bock Nursing Care Beds - Battery
Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.
Identificación
Identificación del producto
- Producto
- Bock Nursing Care Beds - Battery
Información oficial
Riesgo y medidas
- Riesgo o peligro
- In beds with double drives, manufactured in the period from January 2012 to May 2013, some batteries may contain diodes which may be defective. Failure of the diode may lead to damage of the emergency lowering battery. If the bed cannot be lowered this may lead to injury.
- Recomendaciones
- Consumers with impacted stock should contact Alphacare to organise correction. Alphacare will arrange a technician to visit and make the necessary corrections. For further information, contact Alphacare Customer Service on 1300 930 930 or via the Alphacare website at https://alphacare.com.au/contact/
Información oficial
Operadores económicos
- Fabricante
- Alphacare Pty Ltd
Territorio
Geografía
- Distribución
- Retirada en Australia
- País de retirada
- Australia
Procedencia y contexto