OECDReferencia EN/AU/PS1092163
Nimbus 4 and Nimbus Professional Mattresses
Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.
Identificación
Identificación del producto
- Producto
- Nimbus 4 and Nimbus Professional Mattresses
Información oficial
Riesgo y medidas
- Riesgo o peligro
- Nimbus 4 and Nimbus Professional mattresses may over-inflate, due to an inner mattress component failure. This cannot be detected until the product is disassembled. If the mattress over-inflates, patients could fall from the mattress and this could result in serious injury.
- Recomendaciones
- Affected consumers will be contacted in writing. Consumers should follow the instructions provided in the customer letter and the instructional video, to puncture the grommet membrane so that excess air may escape. Removing the membrane will not affect the mattress functionality in any way. For more information, please contact Arjo Australia by phone on 1800 072 040.
Información oficial
Operadores económicos
- Fabricante
- Arjo Australia
Territorio
Geografía
- Distribución
- Retirada en Australia
- País de retirada
- Australia
Procedencia y contexto
NagameAlert distingue el contenido publicado por la fuente de su organización y normalización. Los valores marcados como «NagameAlert» pertenecen a esa capa de tratamiento. Comprueba siempre medidas, alcance y actualizaciones en la publicación oficial.