Saltar al contenido
OECDReferencia EN/AU/PS1092163

Nimbus 4 and Nimbus Professional Mattresses

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Nimbus 4 and Nimbus Professional Mattresses

Información oficial

Riesgo y medidas

Riesgo o peligro
Nimbus 4 and Nimbus Professional mattresses may over-inflate, due to an inner mattress component failure. This cannot be detected until the product is disassembled. If the mattress over-inflates, patients could fall from the mattress and this could result in serious injury.
Recomendaciones
Affected consumers will be contacted in writing. Consumers should follow the instructions provided in the customer letter and the instructional video, to puncture the grommet membrane so that excess air may escape. Removing the membrane will not affect the mattress functionality in any way. For more information, please contact Arjo Australia by phone on 1800 072 040.

Información oficial

Operadores económicos

Fabricante
Arjo Australia

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia