Skip to content
SAFETY GATEReference A12/00131/22

Particulate Respirator — health risk / other

Descriptive title organised by NagameAlert; the official wording is available from the source.

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Particulate Respirator
Brand
UNIMED Union Medikal
Category
Protective equipment
Lot/s
8682960627002 · Produced: 10.2020U25102020
Product domain
Non-foodNagameAlert treatment

Official information

Risk and measures

Risk or hazard
Health risk / other
Official description
The particle/filter retention of the material is insufficient (measured values: as low as 59%). Furthermore, the total filtration capacity of the mask is insufficient (measured mean values: as low as 71%). Consequently, the product does not fulfil the health and safety requirements; thus, even if combined with other recommended measures, it may not properly protect. The product does not comply with the Personal Protective Equipment (PPE) Regulation and with the relevant European standard EN 149.
Measures
Type of economic operator taking notified measure(s): ImporterCategory of measure(s): Withdrawal of the product from the marketDate of entry into force: 25/11/2021

Territory

Geography

Origin
Türkiye
Notifying country or territory
Poland
Distribution
Notificado por Poland

Provenance and context

NagameAlert distinguishes content published by the source from its organisation and normalisation. Values marked “NagameAlert” belong to that processing layer. Always check measures, scope and updates in the official publication.