Skip to content
OECDReference EN/BE/A12/00615/20

Particle filter mask ; Unknown ; Three Dimensional Protective Respiraor (disposable respirator)

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Particle filter mask ; Unknown ; Three Dimensional Protective Respiraor (disposable respirator)
Category
PROTECTIVE_EQUIPMENT
Model
KN95 ; Manufacturing Date: 2020-03-,Testing Date: 2020-03-08 ; Unknown

Official information

Risk and measures

Risk type
true
Risk or hazard
HEALTH_RISK_OTHER ; The particle/filter retention of the material is insufficient (measured value ≤ 86%). Consequently, an excessive amount of particles or microorganisms might pass through the mask, increasing the risk of infection if not combined with additional protective measures. ; The product does not comply with the Personal Protective Equipment Regulation and the relevant European standard EN 149.
Recommendations
MARKING_PRODUCT_WARNINGS
Published priority
true

Territory

Geography

Origin
China (People's Republic of)
Distribution
Retirada en Belgium
Country of recall
Belgium

Provenance and context