700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
700HPPK / 700HPM, High Pressure Pneumatic Calibration manifold, 21 MPA

Official information

Risk and measures

Risk or hazard
The plunger nut threads on the calibration manifold may fail, causing the plunger assembly to be ejected from the manifold at high speed. There is a risk of serious injury to the user if they are struck by the ejected plunger assembly.
Recommendations
Follow this link to register your recalled manifold on the Fluke website. Please include your full contact information, including an email and shipping address. Once registered, you will receive a return postage label so that it can be returned to FLUKE for repair and returned to you free of charge. For more information, contact Fluke on 1300 135 853 (Monday to Friday, 8:30 am – 5:00 pm AEST) or by emailing ordersanz@fluke.com