The Neuro Zti EVO and CLA Cochlear Implants

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
The Neuro Zti EVO and CLA Cochlear Implants

Official information

Risk and measures

Risk or hazard
The implant may stop working due to a complete loss of communication between the implant and the sound processor. If the implant stops working and shuts down and there is a loss of communication between the implant and the sound processor, users will not be able to hear sounds or instructions and this may lead to harm.
Recommendations
Oticon Medical will write to affected consumers to advise them of the issue and provide instructions.   Consumers should be aware of this issue and if they experience a sudden loss of communication, and the problem cannot be isolated to the external components, they should contact their audiologist to make an appointment to have the device checked and repaired.