DS Maref LX9 Sequential Pneumatic Compression Device

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
DS Maref LX9 Sequential Pneumatic Compression Device

Official information

Risk and measures

Risk or hazard
A fault in the programming of the LX9 pumps, may cause users to experience higher pressure in the garments than set on the device. If the pressure in the garments becomes excessive for the user and exceeds the devices pressure specifications, this could result in an increased risk of health complications.
Recommendations
Medi-Rent will update the programming on all pumps which have been programmed with software version 2020.11.01, which will correct the issue. In the interim, consumers are advised to follow the instructions in the customer letter for use of the device until the reprogramming is complete. For further information, contact Medi-Rent by phone on 1300 726 666.

Official information

Economic operators

Manufacturer
Medi-Rent Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia