Philips Respironics - Certain CPAP and Bi-Level PAP Devices

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Philips Respironics - Certain CPAP and Bi-Level PAP Devices

Official information

Risk and measures

Risk or hazard
The polyester-based polyurethane (PE-PUR) sound abatement foam used in Philips Continuous and Non-Continuous Ventilators may: 1. Degrade into particles which may enter the device's air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods, such as ozone; and 2. Emit certain chemicals (volatile organic compounds or VOCs). Foam degradation: The potential risks of particulate exposure include: irritation (skin, eye, and respiratory tract), inflammatory response, headache, asthma, adverse effects to other organs (e.g. kidneys and liver) and toxic carcinogenic effects. VOC emission: The potential risks of chemical exposure due to off-gassing include: headache/dizziness, irritation (eyes, nose, respiratory tract, skin), hypersensitivity, nausea/vomiting, toxic and carcinogenic effects.
Recommendations
It is important that you register your device as soon as possible at the support website www.philips.com/src-update Do not stop or alter your prescribed therapy; stopping treatment suddenly could have an immediate and detrimental effect on your health. Before making any changes to prescribed therapy, make an appointment with your physician or care provider to discuss alternative long term therapy options. Philips will write to affected consumers. Users and carers should review the information provided in the customer letter about the ongoing use of impacted devices. Philips will repair or replace the affected devices. For more information see the TGA safety alert at https://www.tga.gov.au/alert/philips-recall-action-cpap-bi-level-pap-devices-and-mechanical-ventilators For further information, please contact the Philips Support Hotline on 1800 009 579.

Official information

Economic operators

Manufacturer
Philips Electronics Australia Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia