OECDReference EN/AU/PS1091518
Flocare Infinity II Enteral Feeding Pumps
Official information organised for easier consultation. The source publication remains authoritative.
Identification
Product identification
- Product
- Flocare Infinity II Enteral Feeding Pumps
Official information
Risk and measures
- Risk or hazard
- In a situation where the device is not plugged into the main power supply and when running on battery power only, the device may unexpectedly switch off without sounding the alarm to indicate a low battery charge. This issue only affects Flocare Infinity Enteral Feeding Pumps. It does not affect any other Nutricia products. If the device switches off, this could cause a delay in the feeding routine and potentially lead to dehydration and diabetic complications.
- Recommendations
- Nutricia Australia Pty Ltd will contact affected consumers in writing and advise consumers to follow the instructions in the customer letter: 1. Where possible, plug the device into the mains power during feeding therapy (e.g. during overnight feeding). 2. Fully charge the battery before operating the device without mains power (in a mobile use situation). The battery will be completely charged after approximately 6 hours. 3. If users notice that the battery must be charged more frequently than normal, please return the device to the service centre. 4. If the device is running on battery power only and users identify a significant decline in battery capacity or the device is not able to run a normal programmed therapy on a fully charged battery without hearing a low battery charge alarm, please return the device to the service centre. For further information, contact the Nutricia at Home team by phone on 1800 884 367 or the Nutricia Clinical Care Line on 1800 060 051.
Official information
Economic operators
- Manufacturer
- Nutricia Australia Pty Ltd
Territory
Geography
- Distribution
- Retirada en Australia
- Country of recall
- Australia
Provenance and context
NagameAlert distinguishes content published by the source from its organisation and normalisation. Values marked “NagameAlert” belong to that processing layer. Always check measures, scope and updates in the official publication.