Laerdal Compact Suction Unit (LCSU) 4

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Laerdal Compact Suction Unit (LCSU) 4

Official information

Risk and measures

Risk or hazard
The pump of the affected units may malfunction resulting in low suction levels or failure to provide suction. Delay in suctioning secretions, blood or vomit blocking the airways can lead to suffocation if an alternative suction or ventilation method is not administered by the user.
Recommendations
Laerdal will contact affected consumers in writing to arrange for replacement of the affected product. Upon receiving the replacement, consumers should dispose of the pump unit, but keep the carry bag, battery, external power supply and wire stand of the existing unit. Prior to receiving the replacement pump, consumers may continue to use the LCSU4 by following the recommendations below: 1. Do not block the patient tube for more than 10 seconds at a time at vacuum levels greater than 300mmHg; 2. Only use the approved accessories; and 3. Once the desired vacuum level is reached unblock the patient tube immediately. For further information, please contact Laerdal Customer Service by phone on 1800 331 565 (Monday to Friday, 8.30am - 5pm AEST)

Official information

Economic operators

Manufacturer
Laerdal Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia