MiniMed 640G Insulin Infusion Pump

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
MiniMed 640G Insulin Infusion Pump

Official information

Risk and measures

Risk or hazard
Certain pumps are missing solder on specific contacts on the Harness Assembly. If the wires make an electrical contact with the Printed Circuit Board Assembly, the lack of solder could result in increased resistance at the connection point, causing early battery depletion. All necessary battery-related alerts/alarms will still occur to inform the user about the AA battery status. If the battery becomes depleted and the alarm is ignored insulin delivery may be interrupted, this could result in the user developing dangerously high blood glucose levels (hyperglycaemia), which can lead to serious health complications.
Recommendations
Medtronic will provide a replacement pump to all affected consumers. Consumers are advised to record their current pump settings and program the replacement pump provided by Medtronic. Once the replacement pump is successfully programmed and able to be used, consumers are requested to return the affected pump using the shipping materials included with the replacement pump delivery. For further information, please contact Medtronic Customer Service by phone on 1800 777 808.

Official information

Economic operators

Manufacturer
Medtronic Australasia Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia