DDU-2000 Series Automated External Defibrillator (AED)

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
DDU-2000 Series Automated External Defibrillator (AED)

Official information

Risk and measures

Risk or hazard
An electrical component may cause the product to abort a shock delivery, or reset unexpectedly. An aborted shock or unexpected delivery may result in patient death under emergency situations.
Recommendations
Device Technologies Australia will contact impacted customers to organise the return and replacement of affected units. Customers are advised that if alternate units are available for use, please return affected units to Device Technologies and they will be replaced. If no alternate units are available, please inform Device Technologies and continue using the device with caution until replacement is provided. For more information, contact Device Technologies Customer Service on 1300 338 423.

Official information

Economic operators

Manufacturer
Device Technologies Australia Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia