DDU-2000 Series Automated External Defibrillator (AED)
Official information organised for easier consultation. The source publication remains authoritative.
Identification
Product identification
- Product
- DDU-2000 Series Automated External Defibrillator (AED)
Official information
Risk and measures
- Risk or hazard
- An electrical component may cause the product to abort a shock delivery, or reset unexpectedly. An aborted shock or unexpected delivery may result in patient death under emergency situations.
- Recommendations
- Device Technologies Australia will contact impacted customers to organise the return and replacement of affected units. Customers are advised that if alternate units are available for use, please return affected units to Device Technologies and they will be replaced. If no alternate units are available, please inform Device Technologies and continue using the device with caution until replacement is provided. For more information, contact Device Technologies Customer Service on 1300 338 423.