Inogen One G3 Concentrator

Official information organised for easier consultation. The source publication remains authoritative.

Identification

Product identification

Product
Inogen One G3 Concentrator

Official information

Risk and measures

Risk or hazard
When connected to a fully charged 8-cell battery as its only power source, an operating condition may cause an abrupt shut down of the Inogen One G3 Portable Oxygen Concentrator (POC). If no action is taken, the POC may become unresponsive for up to one minute, after which it will behave as though a battery was just connected with a standard flash screen on display. The POC will then be ready for another startup. If powered on again, this cycle may need to be repeated until the 8-cell battery has been adequately discharged allowing the POC to operate as normal. Risk of interruption to the supply of oxygen to users, which may lead to health complications.
Recommendations
Customers are advised that arrangements will be made for the products to be repaired and/or replaced. An Air Liquide Healthcare representative will contact customers via phone to initiate the replacement POC, and collect the affected concentrator. For further information, contact Air Liquide Healthcare Customer Service Centre on 1300 360 202.

Official information

Economic operators

Manufacturer
Air Liquide Healthcare Pty Ltd

Territory

Geography

Distribution
Retirada en Australia
Country of recall
Australia