OECDReferencia EN/AU/PS1104294

DreamWear Full Face Mask, DreamWisp Nasal Mask with Over the Nose Cushion, Wisp Nasal Masks and Wisp Youth Nasal Masks, Amara View Minimal Contact Full-Face Mask and Therapy Mask 3100 NC/SP

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
DreamWear Full Face Mask, DreamWisp Nasal Mask with Over the Nose Cushion, Wisp Nasal Masks and Wisp Youth Nasal Masks, Amara View Minimal Contact Full-Face Mask and Therapy Mask 3100 NC/SP

Información oficial

Riesgo y medidas

Riesgo o peligro
The affected masks contain magnets which can potentially affect the functioning of implanted medical devices such as metal splinters in the eyes, brain shunts, aneurysm clips, pacemakers, implantable cardioverter defibrillators, defibrillators, VP shunts, and neurostimulators used in and around the neck or any implanted medical device impacted by magnetic fields. Philips Respironics is updating the existing contraindications and warnings on the Instructions For Use (IFU) to strengthen the current labelling and warnings related to magnets. If the mask magnets are placed less than 6 inches (approx. 15.24 cm) away from a metallic implant or device,  the magnets may cause the device to not function as intended. This can result  in serious injuries.  Possible impacts may include seizure, arrhythmia, irregular blood pressure, change in heartbeats and cognitive issues.
Recomendaciones
Consumers should read and follow the updated warnings and Instructions For Use (IFU) and take appropriate action: Patients, Household Members, Caregivers and Bed Partners with a medical implant and/or device(s): Must ensure the mask is kept at least 6 inches (approximately 15.24cm) away from the medical implant(s) and/or device(s) Patients should stop using the affected mask if the implant and/or medical device is contraindicated against the mask magnets. Patients should consult their physician immediately to determine if another mask can be used for their therapy and in the interim, switch to a non-magnetic mask if available, for continued therapy. Patients should properly dispose of the mask that has magnets after an alternative is obtained. If patients do not have implanted medical devices, or metallic splinters in their eyes, then no action related to patients is needed. If you are affected or need further information, please contact the Philips Sleep and Respiratory Care hotline on 180 830 517 to organise a suitable remediation which may include a replacement mask or clips to replace the magnets. For more details see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01171-1

Información oficial

Operadores económicos

Fabricante
Philips Electronics Australia Ltd

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia