OECDReferencia EN/AU/PS1104261

Omnipod DASH Personal Diabetes Manager (PDM)

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Omnipod DASH Personal Diabetes Manager (PDM)

Información oficial

Riesgo y medidas

Riesgo o peligro
The product's battery pack may swell, leak fluid or overheat and could catch on fire. If the battery overheats, this could result in a fire which can cause serious injury or death. Additionally If the PDM becomes inoperable there may be a delay in therapy. If insulin delivery is suspended, this may lead to major hyperglycemia which can result in serious health complications. In severe cases of hyperglycemia, the user may experience diabetic ketoacidosis and may require hospitalisation or intervention from a medical professional.
Recomendaciones
Insulet Corporation is implementing a software change to the PDM to limit the charging capacity to mitigate the risk of overcharging. An updated Omnipod DASH PDM will be provided to affected consumers when available. Do not charge the PDM if: i)   The PDM battery compartment is bulging. ii)  The PDM experiences overheating. iii) The PDM does not turn on within 45 minutes of being connected to the charger or turns off immediately after removing from the charger. The PDM can be continued to be used with caution if consumers do not experience the above issues: i)   Do not remove the battery from the PDM for any reason. ii)  Do not continue charging after reaching 85% of charge. Do not charge the device       overnight. iii) Do not charge PDM if surrounding temperature is 31 degree C or higher. Do not charge under direct sunlight. iv) Do not charge the PDM when it feels unusually warm. v)  Do not use PDM while charging. vi) Always inspect PDM before charging. For more details, see https://apps.tga.gov.au/Prod/sara/arn-detail.aspx?k=RC-2022-RN-01321-1 For further information, please contact Insulet Australia Customer Care on 1800 954 074.

Información oficial

Operadores económicos

Fabricante
Insulet Australia Pty Ltd

Territorio

Geografía

Origen
United States
Distribución
Retirada en Australia
País de retirada
Australia