OECDReferencia EN/AU/PS1093286
M8 and S8 Lead Adaptors
Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.
Identificación
Identificación del producto
- Producto
- M8 and S8 Lead Adaptors
Información oficial
Riesgo y medidas
- Riesgo o peligro
- Certain patients who have been implanted with S8 (SADP2008-25(B)) or M8 (MADP2008-25(B)) Lead Adaptors received implant cards (patient ID cards) that incorrectly indicate the system is 'MR Conditional', systems with these lead adaptors should be identified as 'MR Unsafe'. The M8 and S8 Lead Adapters of the implanted Senza system are not MR Conditional. The M8 and S8 Lead Adaptors are correctly noted as MR unsafe in the MRI manual. Magnetic Resonance Imaging (MRI) poses a hazard to consumers who are implanted with S8 or M8 Lead Adaptors. Consumers with 'MR Unsafe' implants should not be scanned. If users are unaware that the lead adaptors are 'MR Unsafe' they may provide incorrect information to medical staff and this may result in a serious injury.
- Recomendaciones
- Emergo Australia will directly contact implanted consumers to provide a new 'MR Unsafe' implant card. For further information, contact Emergo Australia by phone on 02 9006 1662.
Información oficial
Operadores económicos
- Fabricante
- Emergo Asia Pacific Pty Ltd t/as Emergo Australia
Territorio
Geografía
- Distribución
- Retirada en Australia
- País de retirada
- Australia
Procedencia y contexto
NagameAlert distingue el contenido publicado por la fuente de su organización y normalización. Los valores marcados como «NagameAlert» pertenecen a esa capa de tratamiento. Comprueba siempre medidas, alcance y actualizaciones en la publicación oficial.