Saltar al contenido
OECDReferencia EN/AU/PS1093475

Philips Respironics - Certain Ventilator Devices

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Philips Respironics - Certain Ventilator Devices

Información oficial

Riesgo y medidas

Riesgo o peligro
The polyester-based polyurethane (PE-PUR) sound abatement foam used in Philips Continuous Ventilators may: 1. Degrade into particles which may enter the device's air pathway and be ingested or inhaled by the user. The foam degradation may be exacerbated by use of unapproved cleaning methods, such as ozone; and 2. Emit certain chemicals (volatile organic compounds or VOCs) Foam degradation: The potential risks of particulate exposure include: irritation (skin, eye, and respiratory tract), inflammatory response, headache, asthma, adverse effects to other organs (e.g. kidneys and liver) and toxic carcinogenic effects. VOC emission: The potential risks of chemical exposure due to off-gassing include: headache/dizziness, irritation (eyes, nose, respiratory tract, skin), hypersensitivity, nausea/vomiting, toxic and carcinogenic effects.
Recomendaciones
It is important that you register your device as soon as possible at the support website www.philips.com/src-update Do not stop or alter your prescribed ventilator therapy. Make an appointment with your physician to discuss whether it is safe to stop therapy or an alternative ventilator option is required. Phillips will write to affected consumers. Users and carers should review the information provided in the customer letter about the ongoing use of impacted devices. Philips will repair or replace the affected devices. For more information see the TGA safety alert at https://www.tga.gov.au/alert/philips-recall-action-cpap-bi-level-pap-devices-and-mechanical-ventilators For further information, please contact the Philips Support Hotline on 1800 009 579.

Información oficial

Operadores económicos

Fabricante
Philips Electronics Australia Ltd

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia

Procedencia y contexto

NagameAlert distingue el contenido publicado por la fuente de su organización y normalización. Los valores marcados como «NagameAlert» pertenecen a esa capa de tratamiento. Comprueba siempre medidas, alcance y actualizaciones en la publicación oficial.