Dexcom G4 Platinum, G5 Mobile and G6 Sensor Glucose Monitoring System

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Dexcom G4 Platinum, G5 Mobile and G6 Sensor Glucose Monitoring System

Información oficial

Riesgo y medidas

Riesgo o peligro
The Dexcom Continuous Glucose Monitoring System may show falsely elevated glucose readings for users who are taking hydroxyurea, an anti-neoplastic drug, used primarily in chemotherapy. Incorrect glucose readings may result in incorrect diabetes treatment decisions. If incorrect information is used to make treatment decisions, this could lead to an incorrect dose of insulin. This may result in failure to treat low blood glucose (hypoglycemia, loss of consciousness, seizure, death) from missed low sensor glucose values and alerts or by causing other errors (for example, hypoglycemia from an insulin dose based on a falsely elevated sensor glucose reading) and may lead to serious health complications.
Recomendaciones
AA-Med Pty Ltd will contact affected users in writing. When using the G4 Platinum, G5 Mobile or G6 Sensor, consumers should read the additional information in the warning sections of the Instructions For Use (IFU) relating to hydroxyurea. The updated warning is as follows: If you are taking hydroxyurea, your glucose readings may be falsely elevated and result in missed hypoglycemia alerts or errors in diabetes management. The level of inaccuracy depends on the amount of hydroxyurea in your body. Use your meter. For further information, contact Australasian Medical & Scientific Customer Service by phone on 1300 851 056.

Información oficial

Operadores económicos

Fabricante
AA-Med Pty Ltd

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia