Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

Información oficial organizada para facilitar su consulta. La publicación de la fuente conserva la autoridad.

Identificación

Identificación del producto

Producto
Alpha Active 3, Alpha Active 4 and Alpha Active 5 Medical Pump

Información oficial

Riesgo y medidas

Riesgo o peligro
The mains cables may become damaged through incorrect stowage or misuse. If the mains cables have been damaged, this can result in electric shock, which can cause serious injury or death.
Recomendaciones
Arjo Australia will contact affected consumers in writing. This issue will be rectified by increasing the cable wall thickness and incorporating an improved grommet. Consumers should: 1. Verify whether they possess any Alpha Active 3, Alpha Active 4 or Alpha Active 5 pumps and ensure that all caregivers and users are made aware of this issue. 2. Make sure the cable is disconnected from the power socket and then check carefully all electrical connections for signs of damage or excessive wear. If any damage is observed, remove the device from use and contact Arjo Australia. 3. The device can remain in use as long as the cable is not frayed or worn and is secured properly to prevent being caught or entrapped in the bed mechanism as advised in the Instruction for Use. For further information, consumers can contact Arjo Australia on 1800 072 040.

Información oficial

Operadores económicos

Fabricante
Arjo Australia Pty Ltd

Territorio

Geografía

Distribución
Retirada en Australia
País de retirada
Australia